GFM | Gas Flow Meter 2.0

GFM | Gas Flow Meter 2.0
 Description
The Gas Flow Meter 2.0  (GFM) is a portable, explosion-proof, battery-powered direct measurement instrument designed to determine an accurate 
leak rate from various pipe connectors, valves, compressor rod packings/seals and storage tank emissions typically found at compressor stations,
production wells, processing, storage, LNG, regulator and city gate stations in a natural gas environment. 

Fugitive emissions from can be captured by sampling at a high flow rate, where both gas and air pass through the sampler allowing measurement of gas concentration and flow rate. A small-sized Non-Dispersion Infrared (NDIR) optical sensor with a wide temperature range  of –40 to 140°F measures the concentration of natural gas with great accuracy and resolution down to 0.005 SCFM or 0.15 LPM. 

The GFM is installed in a case and can be attached to a backpack harness, leaving the operator’s hands free for use when climbing stairs and descending into confined spaces or use without the backpack harness to hand carry for simple jobs. Use the GFM wirelessly with an Android phone (version 6.0 or higher) to display technical information or assume control of the instrument. Record data by the press of a button as well as a photo of the component in question, if the operator wants to report both across communication channels. Reception distance is up to 15 feet with low energy consumption allowing uninterrupted measurements for approximately 14 hours.

Advantages

Features

Certification

Hazardous Area Intrinsic Safety Evaluation, Testing and Certification is currently underway by UL LLC.

 Specifications

Weight & Dimensions (L x W xH) 9.3 lbs (4.2 kgs):   11.4” x 11.2” x 3.9” (290 x 285 x 100 mm)


Display:                                                       Graphic TFT display - thin film transistor liquid crystal display, active matrix LCD                                                                                       technology, vastly improved image quality and contrast compared to passive matrix LCD


Temperature:                                                 Operational  -40 to 140°F (-20 to 50 °C)


Humidity:                                                     5 to 95% RH (non-condensing)


Battery:                                                       3.7 V Lithium Polymer explosion-proof, low temperature rechargeable.


Capacity:                                                     11.0 Ah


Charging Time:                                              Up to 10 hours


Duration of Work:                                           More than 8 hours (cyclic mode)


Method of Measurement:                                  Pressure drop across the Venturi tube


Measured Quantities:                                       Sample flow rate; background gas concentration; gas concentration in the sample;                                                                                     battery capacity


Estimated Values:                                           Leakage concentration taking into account the background gas level; leak intensity;                                                                                   difference between 1st and 2nd leak rate measurement


Measured Leakage Rate:                                   1.0 to 350.0 l/min


Minimum Detectable Leak Rate:                          From 0.15 l/min


Leak Rate Measurement Error :                           ±5% of reading


Sample Flow Rate:                                           Maximum - 350 l / min; Medium - 250 l / min;


Low - 150 l/min;                                             In two-stage mode second speed lower than initial speed by 25%


Flow Measurement Accuracy:                              ±5% of reading

 